Fluke Corp. MET/CAL Procedure ============================================================================= INSTRUMENT: Omega 869: (1 year) CAL VER /5500 DATE: 26-Dec-94 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 7 NUMBER OF LINES: 76 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # om869.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 30 min. - REQUIRED 1.004 DISP Temperature :18 - 28 degC 1.004 DISP Humidity: <80% 1.006 HEAD INITIAL SETUP 1.007 DISP Remove the back cover. 1.008 PIC om869 1.009 DISP Connect the 5500A to the UUT as follows: 1.009 DISP [32] 5500A NORMAL HI to pin 1 and 4 of probe connector 1.009 DISP [32] 5500A NORMAL LO to pin 2 and 3 of probe connector 1.009 DISP 1.009 DISP Use the same type of wire for all connections. 1.010 HEAD {200degC RANGE} 1.011 DISP Select the 200degC range. 1.012 5500 -100.0degC 0.3U R1 4W 2.001 5500 0.0degC 0.3U R1 4W 3.001 5500 100.0degC 0.3U R1 4W 4.001 5500 195.0degC 0.3U R1 4W 5.001 HEAD {630degC RANGE} 5.002 DISP Select the 630degC range. 5.003 5500 220degC 1U R1 4W 6.001 5500 600degC 1U R1 4W 7.001 END